[1]洪 俊,饶永彩.XN-9000血细胞分析仪筛查外周血中Sezary细胞的临床价值研究[J].现代检验医学杂志,2017,32(03):92-96.[doi:10.3969/j.issn.1671-7414.2017.03.025]
 HONG Jun,RAO Yong-cai.Performance Evaluation of Alarms and Parameters Generated by Hematology Analyzer during Detection of Circulating Sezary Cells of Peripheral Blood Smears[J].Journal of Modern Laboratory Medicine,2017,32(03):92-96.[doi:10.3969/j.issn.1671-7414.2017.03.025]
点击复制

XN-9000血细胞分析仪筛查外周血中Sezary细胞的临床价值研究()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第32卷
期数:
2017年03期
页码:
92-96
栏目:
论著
出版日期:
2017-05-25

文章信息/Info

Title:
Performance Evaluation of Alarms and Parameters Generated by Hematology Analyzer during Detection of Circulating Sezary Cells of Peripheral Blood Smears
文章编号:
1671-7414(2017) 03-092-05
作者:
洪 俊1饶永彩2
武汉大学人民医院检验科,武汉 430060;
2.武汉生物制品所,武汉 4 30060
Author(s):
HONG Jun1RAO Yong-cai2
1.Department of Clinical Laboratory,Renmin Hospital of Wuhan U niversity,Wuhan 430060,China;
2.Wuhan Institute of Biological Products,Wuhan 430060,Ch ina
关键词:
Sezary综合征血细胞分析仪形态学
分类号:
R446.11
DOI:
10.3969/j.issn.1671-7414.2017.03.025
文献标志码:
A
摘要:
目的 评估采用XN-900形态报警和结构参数用于筛检外周 血中Sezary细胞的可行性和有效性。方法 分别采用光学显微镜,XN -9000血细胞分析仪和流式细胞术(FCM)对21例Sezary综合征患者(sezary syndrome,SS)和 三个对照组(皮肤T细胞淋巴瘤对照组48例、慢性淋巴组织增生性B细胞疾病(CLPD)对照组55 例、正常人对照组45例)进行检测,采用对照分析法对Sysmex xn-9000分析生成的淋巴细胞 结构参数(ly-x,ly-y,ly-z)及 “Blast/Abn Lympho(原始/异常淋巴)”报警进行对比 分析,采用受试者工作曲线(ROC)来确定ly-x和ly-z的最佳触发筛查Sezary细胞的临界值 ,对“Blast/Abn Lympho(原始/异常淋巴)”报警提示和结构参数(ly-x和ly-z)的各种组 合筛检规则的诊断效能指标进行评估。结果 Sezary综合征组淋巴细 胞结构参数ly-x和ly-z分别为89.80(87.55~91.52)和73.60(69.10~74.32),明显 高于皮肤T细胞淋巴瘤对照组、CLPD组和正常人对照组(U=37,40,44,55,38,41,均P <0.05);ly-x与Sezary细胞数量正相关,ly-z与大Sezary细胞存在正相关(r=0.248 ,0.198,P<0.05);采用ROC曲线确定ly-x和ly-z筛查Sezary细胞的临界值分别为≥8 6和≥68,将上述三个指标组合起来(即“报警Blast/Abn Lympho≥100,ly-x≥86且ly-z ≥68)的灵敏度和特异度分别为100%和92.1%,可获得最好的筛检Sezary细胞诊断效能。〖H T5”H〗结论 采用血细胞分析仪所产生的“Blast/Abn Lympho(原始/异常淋巴) ” 报警提示和结构参数 (ly-x,ly-z) 组合规则可在外周血中Sezary细胞筛检中获得较 高的特异度和敏感度,可用于疑似SS样本的复片筛检。
Abstract:
Objective To evaluate the performance alarms an d parameters generated by hematology analyzer during detection of circulating se zary cells of peripheral blood smears.Methods Blood samp les from patients with Sezary syndrome (n=21) were studied with Sysmex XN-9 000 analyzer,and compared to patients with benign or tumoral skin lesions (n =48) and patients with chronic lymphoproliferative B-cell diseases (n=55) and normal person (n=45) used as control.Results I n present series,the value of structural lymphoid parameters (ly-x and ly-z) and the alarm Blast/Abn Lympho were statistically higher in Sezary cases than in control cases (U=37,40,44,55,38,41,all P<0.05).Screening Sezary cells critical values of ly-x and ly-z determined by ROC curve were greater than or equal to 86 and was more than or equal to 68.The combination of above t hree indexes up (i.e.“alarm Blast/Abn Lympho>100,ly-x=86 and ly-z>68) sen sitivity and specificity respectively for 100% and 92.1%,in order to get the b est screening Sezary cells diagnostic efficiency.In addition,the value of ly- x was associated to the count of circulating Sezary cells and value of ly-z to the presence of large Sezary cells,both parameters described as prognostic fact ors.Conclusion The combination of alarm Blast/Abn Lympho and structural parameters (ly-x/ly-z/ly-y) may allow to define rule of blood slide review to screen circulating Sezary cells.

参考文献/References:

[1] Swerdlow S,Campo E,Harris N,et al.World Health Organization C lassificationof Tumours of Haematopoietic and Lymphoid Tissues[S].4th Ed.Lyon ,France:IARC,WHO Press:2008.
[2] Talpur R,Singh L,Daulat S,et al.Long-term outcomes of 1 263 patients with mycosis fungoides and Sézary syndrome from 1982 to 2009[J].Clin Cancer Res,2012,18(18):5051-5060.
[3] Devata S,Wilcox RA.Cutaneous T-cell lymphoma:a review with a focus on targeted agents[J].Am J Clin Dermatol,2016,17(3):225-237.
[4] Chhuy J,Morel D,Goeddert G,et al.Pertinence of the sysmexXE-5000 TM parameters:rule of slide review in a context of ‘normal’lymphocyte count (defined from control and mantle cell lymphoma blood specimens)[J].Int J Lab Hematol,2013,35(5):510-516.
[5] Baseggio L,Berger F,Morel D,et al.Identification of circulating CD10 positive T cells in angioimmunoblastic T-cell lymphoma[J].Leukemia,2006,20 (2):296-303.
[6] 洪 俊,饶永彩.应用ROC曲线确定流式细胞术髓过氧化酶检测结果在急性白血病分 类中的最佳临界值[J].现代检验医学杂志,2015,30(2):76-80,83. Hong J,Rao YC.Detecting optimal cut-off value of flow cytometry m yeloperoxidase(MPO) by receiver operator characteristic curve in the classificat ion of acute leukemia[J].J Mod Lab Med,2015,30(2):76-80,83.
[7] Moriarty B, Whittaker S .Diagnosis,prognosis and management of erythro dermic cutaneous T-cell lymphoma[J].Expert Rev Hematol,2015,8(2):159-171 .
[8] Kim YH,Tavallaee M,Sundram U,et al.Phase II investigator-initiated s tudy of brentuximab vedotin in mycosis fungoides and sézary syndrome with varia ble CD30 expression level:a multi-institution collaborative project[J].J Cl in Oncol,2015,33(32):3750-3758.
[9] Horna P,Deaver DM,Qin D,et al.Quantitative flow cytometric identifica tion of aberrant T cell clusters in erythrodermic cutaneous T cell lymphoma impl ications for staging and prognosis[J].J Clin Pathol,2014,67(5):431-436.
[10] Mancebo SE,Cordova M,Myskowski PL,et al.Reflectance confocal microscopy features of mycosis fungoides and Sézary syndrome:correlation with histopathologic and T-cell receptor rearrangement studies[J].J Cutan Pathol ,2016,43(6):505-515.
[11] Sokotowska-WojdytoM,Olek-Hrab K,Ruckemann-Dziurdzińska K.Primar y cutaneous lymphomas: diagnosis and treatment[J].Postepy Dermatol Alergol ,2015,32(5):368-383.

相似文献/References:

[1]孟 欣,王 刚,袁 莉,等.Sysmex血细胞分析仪白细胞分类计数校准方法的建立与验证[J].现代检验医学杂志,2016,31(01):101.[doi:10.3969/j.issn.1671-7414.2016.01.029]
 MENG Xin,WANG Gang,YUAN Li,et al.Establishment of Method and Verification for Leukocyte Differential Count of Sysmex Hematology Analyzer[J].Journal of Modern Laboratory Medicine,2016,31(03):101.[doi:10.3969/j.issn.1671-7414.2016.01.029]
[2]张金彪,代荣琴.ISO 15189 医学实验室认可标准下血细胞分析仪故障修复后的检验质量验证[J].现代检验医学杂志,2020,35(05):168.[doi:10.3969/j.issn.1671-7414.2020.05.043]
 ZHANG Jin-biao,DAI Rong-qin.Verification of Inspection Quality after Repair of Blood Cell Analyzer Failure by ISO 15189 Medical Laboratory Accreditation Standards[J].Journal of Modern Laboratory Medicine,2020,35(03):168.[doi:10.3969/j.issn.1671-7414.2020.05.043]

备注/Memo

备注/Memo:

作者简介:洪 俊(1973-),男,博士,副主任技师,主要从事临 床免疫及血液学方面的研究,E-mail:wonderman1973@hotmail.com。