[1]肖春海,梁 爽,黄敏洁,等.罗氏U701和希森美康UF-1000i两种尿沉渣分析仪不同分析原理检测细菌结果差异性比较[J].现代检验医学杂志,2019,34(04):132-134.[doi:10.3969/j.issn.1671-7414.2019.04.032]
 XIAO Chun-hai,LIANG Shuang,HUANG Min-jie,et al.Comparison of Bacterial Detection Desults between Roche U701 andSysmex UF-1000i Urine Sediment Analyzer Based on Different Analysis Principles[J].Journal of Modern Laboratory Medicine,2019,34(04):132-134.[doi:10.3969/j.issn.1671-7414.2019.04.032]
点击复制

罗氏U701和希森美康UF-1000i两种尿沉渣分析仪不同分析原理检测细菌结果差异性比较()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第34卷
期数:
2019年04期
页码:
132-134
栏目:
研究简报·实验技术
出版日期:
2019-08-20

文章信息/Info

Title:
Comparison of Bacterial Detection Desults between Roche U701 andSysmex UF-1000i Urine Sediment Analyzer Based on Different Analysis Principles
文章编号:
1671-7414(2019)04-132-04
作者:
肖春海梁 爽黄敏洁董志武
(上海市第六人民医院金山分院检验科,上海 201500)
Author(s):
XIAO Chun-haiLIANG ShuangHUANG Min-jieDONG Zhi-wu
(Department of Clinical Laboratory,Jinshan Branch of Shanghai Sixth People's Hospital,Shanghai201500,China)
关键词:
尿有形成分分析 尿液细菌培养 尿路感染
分类号:
R446
DOI:
10.3969/j.issn.1671-7414.2019.04.032
文献标志码:
A
摘要:
目的 以中段尿细菌培养为标准,比较荧光染色法和图像智能判读法两种原理的尿沉渣分析仪细菌结果的差异性。方法 对选取上海市第六人民医院金山分院2018年11月疑似尿路感染患者的223份中段尿样品进行细菌培养和鉴定,接种之后,该样品分别在罗氏U701和希森美康UF-1000i上进行检测,最后对尿标本进行离心涂片染色镜检。以培养结果为金标准,分析两台设备的敏感度、特异度和正确度,从而比较两台设备检测细菌的性能。结果 223份尿样品培养出革兰阳性菌20株,革兰阴性菌35株。UF-1000i,离心涂片染色镜检和U701的结果分别与细菌培养结果比较,UF-1000i和离心涂片染色镜检的正确度分别为98.2%和99.1%,且差异无统计学意义(均P>0.05)。U701的三种参考范围的结果中,“-”组、“-~1+”和“-~2+”的正确度分别为67.3%,91.5%和87.4%,与细菌培养结果比较差异均有统计学意义(均P<0.05)。结论 UF-1000i的细菌结果对于尿路感染的提示优于U701,U701细菌结果的参考范围建议为“-~1+”。
Abstract:
Objective The middle-stage urine bacterial culture as the standard, to compare the bacterial results of the two urine sediment analyzers with different principles of fluorescence staining and image intelligent interpretation.Methods 223 mid-stream specimens ofurine in Jinshan Branch of the Shanghai Sixth People's Hospital in November 2018 were cultivated and identified.After the culture,the specimens were testedby Roche U701 and Sysmex UF-1000i respectively.The culture result was as the gold standard.The sensitivity,specificity and coincidence rate of the two devices were analyzed.Then,it was compared that the performance of the bacterial results of the two devices.Results 20 strains of Gram-positive bacteria and 35 strains of Gram-negative bacteria were cultivated in 223urine specimens.The results of UF-1000i,centrifugation staining microscopy and U701 were compared with those of bacterial culture.The coincidence rate of UF-1000i and centrifugal staining microscopy was 98.2% and 99.1%,respectively.And no statistical difference(all P>0.05).The results of the three reference ranges of U701,the coincidence rates of “-” group,“-~1+” and “-~2+” were 67.3%,91.5% and 87.4%,respectively.The results of the three grouping methods of U701 were statistically different from the results of bacterial culture(all P<0.05).Conclusion UF-1000i were betterthan U701 for urinary tract infection in the bacterial results.The reference range for U701 bacterial result was recommended as “-~1+”.

参考文献/References:

[1] 严海忠,王伟佳,慕月晶,等.血液及尿液中降钙素原(PCT)检测在泌尿系统感染定位中的诊断价值[J].现代检验医学杂志,2017,32(4):116-118. YAN Haizhong,WANG Weijia,MU Yuejing,et al.Diagnosticv value of detecting the serum and urine procalcitonin(PCT)for the urinary system infection[J].Journal of Modern Laboratory Medicine,2017,32(4):116-118. [2] 刘思娣,黄勋,曾翠,等.不同类别重症监护病房持续3年医院感染前瞻性目标性监测[J].中国感染控制杂志,2019,18(1):17-21. LIU Sidi,HUANG Xun,ZENG Cui,et al.Prospective targeted surveillance on healthcare-associated infection in different types of intensive care units for three consecutive years[J].Chinese Journal of Infection Control,2019,18(1):17-21. [3] SEIFU W D,GEBISSA A D.Prevalence and antibiotic susceptibility of uropathogens from cases of urinary tract infections(UTI)in Shashemene referral hospital,Ethiopia:[J].BMC Infectious Diseases,2018,18(1):30. [4] 田鹏鹏,朱丽莎,马青,等.尿路感染中肠球菌的分布及耐药性分析[J].中华医院感染学杂志,2017,27(4):743-745,761. TIAN Pengpeng,ZHU Lisha,MA Qing,et al.Distribution and drug resistance of Enterococcus in urinary tract infection[J].Chinese Journal of Nosocomiology,2017,27(4):743-745,761. [5] 梅方超,戴海英,尚小玲,等.尿液HBP,LE,U-NIT和WBC联合检测与尿液细菌定量培养对急性尿道感染的诊断意义[J].现代检验医学杂志,2018,33(5):126-128. MEI Fangchao,DAI Haiying,SHANG Xiaoling,et al.Diagnostic significance of urine HBP,LE,U-NIT and WBC combined with urine bacterial quantitative culture in the diagnosis of acute urethral infection[J].Journal of Modern Laboratory Medicine,2018,33(5):126-128. [6] 阎平.尿路感染诊断中尿常规检验的临床价值与结果分析[J].中国实用医药,2018,13(2):38-39. YAN Ping.Clinical value and result analysis of urinary routine examination in diagnosis of urinary tract infection[J].China Practical Medical,2018,13(2):38-39. [7] 虞培娟,张杲琳,严茹红,等.尿液有形成分分析联合血清降钙素原检测在尿路感染诊断中的应用价值[J].检验医学,2018,33(4):299-304.YU Peijuan,ZHANG Gaolin,YAN Ruhong,et al.Urinary sediment analysis combined with serum procalcitonin determination in the diagnosis ofurinary tract infection[J].Laboratory Medicine,2018,33(4):299-304. [8] 刘小宁.尿常规亚硝酸盐检测对细菌性尿路感染的诊断价值研究[J].中国医药指南,2018,16(7):97. LIU Xiaoning.Diagnostic value of urine routine nitrite detection for bacterial urinary tract infection[J].Guide of China Medicine,2018,16(7):97. [9] 刘宝芹.尿液分析仪检测结果的影响因素及对策[J].医疗装备,2018,31(5):203-204. LIU Baoqin.Influencing factors of urine analyzer test results andcountermeasures[J].Medical Equipment,2018,31(5):203-204. [10] 王东.尿液试纸暴露时间对尿液分析仪检测结果的影响[J].医疗装备,2018,31(12):49-50. WANG Dong.Effects of exposure time of urine dipsticks on urine analyzer results[J].Medical Equipment,2018,31(12):49-50. [11] 冯玉青,黄素素,罗靖,等,UF-1000i全自动尿液分析仪的细菌通道检测与中段尿培养的正确度分析[J].检验医学,2018,33(1):60-62. FENG Yuqing,HUANG Susu,LUO Jing,et al.Co-nsistency rate of UF-1000iautomatic urine analyzer bacterial channel and mid-stream urine culturing[J].Laboratory Medicine,2018,33(1):60-62. [12] CONKAR S,MIR S.Urine flow cytometry in the diagnosis of urinary tractinfection[J].The Indian Journal of Pediatrics,2018,85(11):995-999. [13] MONSEN T,RYDEN P.A new concept and a comprehensive evaluation of SYSMEX UF-1000i flow cytometer to identify culture-negative urine specimens in patients with UTI[J].European Journal of Clinical Microbiology & Infectious Diseases,2017,36(9):1691-1703. [14] 曾仲麟,林雅媛,李健茹,等.UF-1000i尿沉渣分析仪在尿路感染诊断的评价[J].实验与检验医学,2017,35(3):420-422. ZENG Zhonglin,LIN Yayuan,LI Jianru,et al.Evaluation of UF-1000i urinary sediment analyzer in diagnosis of urinary tract infection[J].Experimentaland Laboratory Medicine,2017,35(3):420-422. [15] 刘肖肖,孔德友,刘广宇,等.基于融合算法的尿沉渣显微图像有形成分外边缘检测方法研究[J].中国医学装备,2018,15(11):28-31. LIU Xiaoxiao,KONG Deyou,LIU Gusngyu,et al.Research on the detection method of externsl edge of morphological components of microscopic images of urinesediment based fusion algorithms[J].China Medical Equipment,2018,15(11):28-31.

备注/Memo

备注/Memo:
基金项目:上海市第六人民医院联合体课题(编号:2017-01)。 作者简介:肖春海(1976-),男,硕士,副主任技师,主要从事临床检验与实验室管理,E-mail; xchwjf@163.com。收稿日期:2019-04-04 修回日期:2019-04-29