[1]毛小欢,郭书忍,李 刚.临床POCT 血糖仪质量控制的实践与探讨[J].现代检验医学杂志,2020,35(04):162-164.[doi:10.3969/j.issn.1671-7414.2020.04.041]
 MAO Xiao-huan,GUO Shu-ren,LI Gang.Practice and Discussion on the Quality Control of Hospital POCT Blood Glucose Meters[J].Journal of Modern Laboratory Medicine,2020,35(04):162-164.[doi:10.3969/j.issn.1671-7414.2020.04.041]
点击复制

临床POCT 血糖仪质量控制的实践与探讨()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第35卷
期数:
2020年04期
页码:
162-164
栏目:
质量控制 实验室管理
出版日期:
2020-09-05

文章信息/Info

Title:
Practice and Discussion on the Quality Control of Hospital POCT Blood Glucose Meters
文章编号:
1671-7414(2020)04-162-03
作者:
毛小欢1郭书忍2李 刚1
(1. 河南省人民医院检验科,郑州 450003;2. 郑州大学第一附属医院检验科,郑州 450002)
Author(s):
MAO Xiao-huan1 GUO Shu-ren2 LI Gang1
(1.Department of Clinical Laboratory, Henan Provincial People’s Hospital, Zhengzhou 450003, China; 2. Department of Clinical Laboratory, the First Affiliated Hospital of Zhengzhou University, Zhengzhou 450002, China)
关键词:
床旁即时检测血糖仪质量管理
分类号:
R446
DOI:
10.3969/j.issn.1671-7414.2020.04.041
文献标志码:
A
摘要:
目的 通过对临床床旁即时检测(point-of care testing, POCT)血糖仪进行比对,探讨临床POCT 血糖仪的管理办法。方法 选取136 台临床POCT 血糖仪,采用毛细血管血与静脉血比对方案,以全自动生化分析仪测定葡萄糖为参考值,计算每一台血糖仪在不同浓度水平下测定血糖的偏倚;按血糖仪型号分为三个组,比较各型号血糖仪的偏倚分布情况是否具有统计学意义。结果 符合要求的血糖仪112 台,占比82.35%;不符合要求的血糖仪24 台,占比17.65%;112 台符合要求的血糖仪在血糖浓度低于4.2mmol/L 时,偏倚为-0.683~0.475mmol/L;当血糖浓度介于6.7~11.1mmol/L 之间时,偏倚为-13.26%~17.55%;当血糖浓度介于11.1~16.6mmol/L 之间时,偏倚为-10.90% ~ 13.98%;三种型号的血糖仪与全自动生化分析仪测定结果的偏倚分布情况,差异无统计学意义。三个浓度范围下的P 值分别为0.754,0.587 和0.486。结论 112 台POCT 血糖仪符合比对要求,三种型号的POCT 血糖仪偏倚情况的差异无统计学意义。

参考文献/References:

[1] 宋蓓, 周志男, 王丹, 等. 医院内POCT 血糖室间质量评价方法的建立[J]. 国际检验医学杂志,2019,40(18):2298-2301. SONG Bei, ZHOU Zhinan, WANG Dan, et al. Establishmentof a method for evaluating the quality ofPOCT blood glucose in the hospital [J]. InternationalJournal of Laboratory Medicine, 2019, 40(18): 2298-2301.
[2] 余珍.基层医院POCT 血糖仪室内质控管理现状分析与改进效果评价[C]. 上海:第四届上海国际护理大会论文汇编,2019:210. YU Zhen. Analysis of current situation and evaluationof improvement effect of POCT blood glucose meterindoor quality control management in primary hospitals[C]. Shanghai :Proceedings of the 4th ShanghaiInternational Nursing Conference, 2019:210.
[3] 中华人民共和国卫生部办公厅. 医疗机构便携式血糖检测仪管理和临床操作规范(试行)[Z]. 北京:卫办医政发[2010]209 号,2011. General Office of the Ministry of Health, PRC.Regulations on management and clinical operation ofportable blood glucose detector in medical institutions(Trial) [Z]. Beijing: The Medical Administration ofthe Public Health Office[2010] No 209 ,2011.
[4] National Committee for Clinical Laboratory Standards.Ancillary (bedside) blood glucose testing in acute and chronic care facilites, approved guideline C30-A [S].Villanov: NCCLS, 2002.
[5] 陈娟, 颜巍, 周竞, 等.POCT 血糖仪和生化分析仪对血糖检测的比对研究[J]. 检验医学与临床,2019,16(2):171-173. CHEN Juan,YAN Wei,ZHOU Jing,et al . Thecomparison of blood glucose level detection betweenPOCT blood glucose meter and biochemical analyzer[J]. Laboratory medicine and clinical, 2019, 16 (2):171-173.
[6] 李广权, 周卫东, 钟培英, 等. 不同浓度氟化钠抗凝血对POCT 血糖仪校准结果干扰影响的研究[J]. 现代检验医学杂志,2019,34(1):127-129,132. LI Guangquan, ZHOU Weidong, ZHONG Peiying, etal. Anticoagulant blood of different sodium fluorideconcentration interfered calibrated results of POCTglucometer [J].Journal of Modern Laboratory Medicine,2019,34(1):127-129,132.
[7] 王晟,段小勇,刘兴态,等. 临床POCT 检测项目质量管理体系的建立与实践[J]. 现代检验医学杂志,2019,34(2):147-152. WANG Sheng, DUAN Xiaoyong, LIU Xingtai, et al.Establishment and practiceof quality managementsystem for clinical POCT tesing project [J]. Journalof Modern Laboratory Medicine, 2019, 34(2):147-152.
[8] 何亚妮, 李海炜, 伍惠玲, 等. 88 台4 种型号POCT 血糖仪检测血糖结果的探讨[J].广西医科大学学报, 2012,29(1):85-86. HE Yani, LI Haiwei, WU Huiling, et al. A study on theblood glucose results of 88 POCT blood glucose meterswith 4 types[J]. Journal of Guangxi Medical University,2012, 29(1): 85-86.

相似文献/References:

[1]汪怀周,陈 燕,岳展伊,等.C反应蛋白测定两种检测系统间结果的比较[J].现代检验医学杂志,2016,31(03):127.[doi:10.3969/j.issn.1671-7414.2016.03.036]
 WANG Huai-zhou,CHEN Yan,YUE Zhan-yi,et al.Comparison of Different Assay System for C-reactive Protein Detection[J].Journal of Modern Laboratory Medicine,2016,31(04):127.[doi:10.3969/j.issn.1671-7414.2016.03.036]

备注/Memo

备注/Memo:
收稿日期:2020-03-17 修回日期:2020-05-17作者简介:毛小欢(1982-),男,硕士,主管检验师,主要从事实验室质量管理和分子生物学检验工作。通讯作者:李刚,副主任技师,主要从事实验室管理及信息化研究,E-mail:ligang65587359@163.com。
更新日期/Last Update: 2020-10-08