[1]张洪波,张 辉,黄 铭,等.基于UF1000i 尿沉渣分析仪对Clinitek Atlas 尿干化学分析仪白细胞计数与反射率的关系研究及思考[J].现代检验医学杂志,2022,37(04):165-168.[doi:10.3969/j.issn.1671-7414.2022.04.032]
 ZHANG Hong-bo,ZHANG Hui,HUANG Ming,et al.Research and Thinking on the Relationship between White Blood Cell Count and Reflectance of Clinitek Atlas Urine Dipstick Chemistry Analyzer Based on UF1000i Urine Sediment Analyzer[J].Journal of Modern Laboratory Medicine,2022,37(04):165-168.[doi:10.3969/j.issn.1671-7414.2022.04.032]
点击复制

基于UF1000i 尿沉渣分析仪对Clinitek Atlas 尿干化学分析仪白细胞计数与反射率的关系研究及思考()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第37卷
期数:
2022年04期
页码:
165-168
栏目:
研究简报·实验技术
出版日期:
2022-07-15

文章信息/Info

Title:
Research and Thinking on the Relationship between White Blood Cell Count and Reflectance of Clinitek Atlas Urine Dipstick Chemistry Analyzer Based on UF1000i Urine Sediment Analyzer
文章编号:
1671-7414(2022)04-165-04
作者:
张洪波 张 辉 黄 铭 陈 星李 果
华中科技大学同济医学院附属同济医院检验科,武汉 430030
Author(s):
ZHANG Hong-bo ZHANG Hui HUANG Ming CHEN Xing LI Guo
Department of Laboratory Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430030,China
关键词:
尿干化学分析仪反射率中性粒细胞脂肪酶白细胞脂肪酶校准验证
分类号:
R446.12
DOI:
10.3969/j.issn.1671-7414.2022.04.032
文献标志码:
A
摘要:
目的 借助UF1000i 尿沉渣分析仪,验证尿干化学分析仪中白细胞计数和反射率之间的潜在关系,从而为尿干化学分析仪的定量改进提供新的思路。方法 通过筛选尿白细胞计数大于1 000/μl 的尿标本,在UF1000i 尿沉渣分析仪和Clinitek Atlas 尿干化学分析仪上同时测试,记录尿沉渣白细胞计数和尿干化学反射率值,并统计分析两者的相关定性等级的反射率临界点和白细胞临界点验证。结果 尿沉渣白细胞计数(X)和反射率(Y)存在较好的关系曲线:Y =109.14 ln(X)+931.94,r=0.995 8;尿干化学微量、1+,2+ 和3+ 的临界值反射率分别为1 150,1 231,1 380 和1 553;对应的白细胞临界值(个/μl)分别为5,10,52 和330。结论 尿干化学白细胞计数与反射率的确立,为现行的尿干化学分析仪的校准提供了新的解决思路,也为尿干化学分析仪由定性向定量发展奠定了基础,从而使尿干化学检测服务于临床的质量更进一步。
Abstract:
Objective To verify the potential relationship between the white blood cell(WBC) count and reflectance in urine dipstick chemistry analyzer by means of UF1000i urine sediment analyzer, so as to provide a new idea for quantitative improvement of urine dipstick chemstry analyzer. Methods By screening suitable urine samples(WBC count >1 000/μl) and testing them simultaneously on UF1000i urine sediment analyzer and Clinitek Atlas urine chemical analyzer, the WBC count and urine dipstick chemistry reflectance of urine sediment were recorded, and the correlation between them was calculated. Results The curve between the urine WBC count and reflectance was good: Y = 109.14 ln (X) + 931.94 , and the value of r was 0.995 8.The critical reflectance of trace, 1+, 2+ and 3+ were 1 150,1 231,1 380 and 1 553, respectively .The corresponding WBC count (/μl) were 5,10,52 and 330,respectively. Conclusion The establishment of the WBC count and reflectivity of the urine dipstick chemical analyzer provides a new solution for the calibration of current urine dipstick chemistry analyzer, and also lays a foundation for the development of urine dipstick chemistry analyzer from qualitative to quantitative, so as to further improve the clinical quality of urine dipstick chemistry test.

参考文献/References:

[1] OKUBO S, YOKOTA H, IKEDA H, et al. Overview of ISO15189 and accreditation with it[J]. Rinsho Byori,2010, 58(7): 676-684.
[2] KUBONO K. Quality management system in the medical laboratory-ISO15189 and laboratory accreditation[J]. Rinsho Byori, 2004, 52(3): 274-278.
[3] KILLEEN A A, LONG T, SOUERS R, et al. Verifying performance characteristics of quantitative analytical systems: calibration verification, linearity, and analytical measurement range[J]. Archives of Pathology & Laboratory Medicine, 2014, 138(9): 1173-1181.
[4] DOLSCHEID-POMMERICH R C, KLARMANNSCHULZ U, CONRAD R, et al. Evaluation of the appropriate time period between sampling and analyzing for automated urinalysis[J]. Biochemia Medica, 2016, 26(1): 82-89.
[5] Clinical Laboratory Standards Institute(CLSI). CLSI document GP16-A3:Urinalysis; approved guideline-third Edition [s]. Wayne: PA, CLSI GP16-A3, 2009.
[6] 张洪波, 张驰, 李果. 基于UF1000i 尿沉渣分析仪的ATLAS 尿干化学分析仪红细胞反射率验证思考[J].检验医学, 2017, 32(7):614-615. ZHANG Hongbo, ZHANG Chi, LI Guo. Verification of red blood cell reflectance by ATLAS urinary chemistry dipstick analyzer based on UF1000i urinary sediment analyzer[J]. Laboratory Medicine, 2017, 32(7): 614-615.
[7] YANG Wei, YU Xiaochen, LIU Dan, et al. High false positive rate of white blood cells in urine samples of pregnant women may be caused by epithelial cells being misclassified by the Sysmex UF-1000i urine flow cytometer[J]. Cytometry Part B-Clinical Cytometry,2019, 96(6): 464-468.
[8] SWAMYDAS M, LUO Yi, DORF M E, et al.Isolation of mouse neutrophils[J]. Current Protocols in Immunology, 2015, 110(20): 3.20.1-3.20.15.
[9] HU Yan. Isolation of human and mouse neutrophils ex vivo and in vitro[J]. Methods in Molecular Biology(Clifton, N.J.), 2012, 844: 101-113.
[10] WANG Lina, WANG Huacheng, ZHAO Chanjing,et al. The standardization of the report for urine cell counting-A converting factor for Sysmex UF-1000i[J].Journal of Clinical Laboratory Analysis, 2019, 33(4):e22857.
[11] FABBRO C, DAROLLES J, RAULT J P. Evaluation of the performances of the UF-1000i? automated urine analyzer[J]. Annales de Biologie Clinique, 2011, 69(4):431-439.
[12] LEE W, HA J S, RYOO N H. Comparison of the automated Cobas U 701 urine microscopy and UF-1000i flow cytometry systems and manual microscopy in the examination of urine sediments[J]. Journal of Clinical Laboratory Analysis, 2016, 30(5): 663-671.
[13] MANONI F, TINELLO A, FORNASIERO L, et al.Urine particle evaluation: a comparison between the UF-1000i and quantitative microscopy[J]. Clinical Chemistry and Laboratory Medicine, 2010, 48(8):1107-1111.
[14] 中华人民共和国医药行业标准.YY/T 0475-2011: 干化学尿液分析仪[S]. 北京:国家食品药品监督管理局, 2011. Industry Standard of the People’s Republic of China.YY/T 0475-2011:Dry chemistry urine analyzer[S]. Beijing:National Medical Products Administration,2011.
[15] 中华人民共和国国家计量技术规范.JJF 1129-2005:尿液分析仪校准规范[S]. 北京: 国家质量监督检验检疫总局,2005. National Metrological Technical Specification of the People’s Repulic of China. JF 1129-2005:Calibration specification of urine analyzers[S]. Beijing: General Administration of Quality Supervision, Inspection and Quarantine, 2005.
[16] PENDERS J, FIERS T, DELANGHE J R. Quantitative evaluation of urinalysis test strips[J]. Clinical Chemistry, 2002, 48(12): 2236-2241.
[17] OYAERT M, DELANGHE J. Progress in automated urinalysis[J]. Annals of Laboratory Medicine, 2019,39(1): 15-22.
[18] PIECH T L, WYCISLO K L. Importance of urinalysis[J]. The Veterinary Clinics of North America. Small Animal Practice, 2019, 49(2): 233-245.

备注/Memo

备注/Memo:
作者简介:张洪波(1978-),男,硕士,主管技师,主要从事临床检验工作,E-mail:58639096@qq.com。
通讯作者:李果,Tel:027-83663185。