[1]龙一飞,林冬玲,陈 富,等.基于EP12-A2文件评价尿蛋白定性检测性能[J].现代检验医学杂志,2015,30(02):117-119,122.[doi:10.3969/j.issn.1671-7414.2015.02.036]
 LONG Yi-fei,LIN Dong-ling,CHEN Fu,et al.Evaluate the Performance of Urine Protein Qualitative Test by EP12-A2 Document[J].Journal of Modern Laboratory Medicine,2015,30(02):117-119,122.[doi:10.3969/j.issn.1671-7414.2015.02.036]
点击复制

基于EP12-A2文件评价尿蛋白定性检测性能()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第30卷
期数:
2015年02期
页码:
117-119,122
栏目:
研究简报·实验技术
出版日期:
2015-03-20

文章信息/Info

Title:
Evaluate the Performance of Urine Protein Qualitative Test by EP12-A2 Document
作者:
龙一飞林冬玲陈 富许振杰陈中华刘建平李 松李有强王丽娜
广东省中医院检验科,广州 510000
Author(s):
LONG Yi-feiLIN Dong-lingCHEN FuXU Zhen-jieCHEN Zhong-huaLIU Jian-pingLI Song LI You-qiangWANG Li-na
Department of Clinical Laboratory, Guangdong Provincial Hospital of Chinese Medicine,Guangzhou 510000,China
关键词:
尿蛋白定性 性能评价 EP12-A2
分类号:
R446.112
DOI:
10.3969/j.issn.1671-7414.2015.02.036
文献标志码:
A
摘要:
目的 评价AX-4030与Cobas U411尿干化学分析仪在尿蛋白定性检测的分析性能,探讨定性试验性能评价方法。方法 参照美国临床实验室标准化协会(CLSI)EP12-A2文件推荐方法,对日本京都AX-4030和罗氏Cobas U411尿干化学分析仪两检测系统尿蛋白定性检测做偏倚和不精密度分析,比较两检测系统的C50, C5~C95区间以及不精密度区间。两系统同时检测310份临床尿标本蛋白,统计分析与评价两者结果的一致程度。结果 尿干化学分析仪AX-4030检测系统检测尿蛋白的C50小于CobasU411系统,AX-4030系统的C5~C95区间比U411系统的窄,AX-4030系统的不精密度曲线较U411系统曲线陡峭; 两套系统检测临床标本尿蛋白结果方法学的比较,一致性百分率为96.8%,阳性一致率为82.7%,阴性一致率为99.6%; 95%CI为94.2%~98.16%,Kappa值为0.88。结论 AX-4030检测系统在尿蛋白C50临界值处的灵敏度及不精密度较Cobas U411系统稍优,二者在临床标本检测中的尿蛋白结果一致性程度极强。EP12-A2推荐的定性试验性能评价方法可行、有效。
Abstract:
Objective To evaluate the analysis capability of urine protein qualitative test between AX-4030 and Cobas U411 urine dry chemistry analyzer,and study on evaluating the performance of qualitative test.Methods According to Clinical and Laboratory Standards Institute(CLSI)EP12-A2 document,analyzed the bias and imprecision of urineprotein qualitative test between the Aution MAX AX-4030 and Roche CobasU411 system.Their C50,C5~C95 intervals and imprecision curves were compared.Theprotein of 310 specimens were simultaneously determined by both Cobas U411 andAX-4030,in order to evaluate their concordance.Results C50 for AX-4030 system was less than that for Cobas U411; C5~C95 interval of AX-4030 system was narrower than CobasU411.The imprecision curve of AX-4030 system was steeper than Cobas U411.The comparison of the two analysis systems showed that the concordance was 96.8%,the positive concordance was 82.7%,and the negative concordance was 99.6%.The 95% credibility interval(CI)was 94.2%~98.16% and the Kappa value was 0.88.ConclusionFor the sensitivity and imprecision of urine protein test in the C50 critical value,the AX-4030 system was better than Cobas U411.The concordance of them in determining clinical specimens was pole-strength.The evaluation recommended by the EP12-A2 document is practical and effective.

参考文献/References:

[1] Clinical and Laboratory standards Institute.User protocol for evaluation of qualitative test performance; approved guideline[S].Wayne:PA,GLSI EP12-A2,2008.
[2] 中国合格评定国家认可委员会.医学实验室质量和能力认可准则[S].北京:CNAS-CL02,2008. China National Accreditation Service for Co-nformity Assessment.Accreditation criteria for the quality and competence of medical Laboratories[J].Beijing:CNAS-CL02,2008.
[3] 陈 燕,赵 敏,张家红,等.尿微量蛋白检查对糖尿病早期肾损伤诊断价值[J].中华检验医学杂志,2003,26(9):562-564. Chen Y,Zhao M,Zhang JH,et al.Monitoring diabetes nephropathy byurine examination[J].Chinese Journal of Laboratory Medicine,2003,26(9):562-564.
[4] 夏邦世,吴金华.Kappa一致性检验在检验医学研究中的应用[J].中华检验医学杂志,2006,29(1):83-84. Xia BS,Wu JH.Kappa consistency inspection in the application of medical research[J].Chinese Journal of Laboratory Medicine,2006,29(1):83-84.
[5] 陈桂山,张秀明,熊继红,等.未知诊断定性试验分析性能评价方法探讨[J].检验医学,2010,25(12):978-981. Chen GS,Zhang XM,Xiong JH,et al.Study on evaluating the performance of unknown diagnosis qualitative test[J].Laboratory Medicine,2010,25(12):978-981.
[6] 王洁贞,韩 兢,刘言训,等.Kappa统计量在一致性和重现性检验中的应用[J].山东医科大学学报,1996,34(3):209-212. Wang JZ,Han J,Liu YX,et al.Kappa statistics in the application of the consistency and reproducibility test[J].Journal of Shandong Medical University,1996,34(3):209-212.
[7] Landis JR,Koch GG.The measurement of observer agreement for Gategoricaldate[J].Biometrics,1977,33:159-174.

相似文献/References:

[1]肖亚玲,吴皓瑜,王 琪,等.应用临床检验定量测定项目室内质量控制数据监测平台对室内质量控制实验室间比对数据的分析[J].现代检验医学杂志,2016,31(02):149.[doi:10.3969/j.issn.1671-7414.2016.02.045]
 XIAO Ya-ling,WU Hao-yu,WANG Qi,et al.Application of Internal Quality Control Data Monitor Platform for Quantitative Testing in Clinical Laboratory for the Analysis of Inter-Laboratory Comparison[J].Journal of Modern Laboratory Medicine,2016,31(02):149.[doi:10.3969/j.issn.1671-7414.2016.02.045]
[2]汪靖园,王林川,肖 尧,等.科华ELISA与雅培i2000sr检测系统对4种感染性疾病8项血清标志物检测的性能评价[J].现代检验医学杂志,2018,33(06):125.[doi:10.3969/j.issn.1671-7414.2018.06.033]
 WANG Jing-yuan,WANG Lin-chuan,XIAO Yao,et al.Performance Evaluation of Kehua ELISA and Abbott i2000sr in Eight Serum Markers Detection for Four Infectious Diseases[J].Journal of Modern Laboratory Medicine,2018,33(02):125.[doi:10.3969/j.issn.1671-7414.2018.06.033]
[3]方佩琪,孙 林,顾梅秀,等.碧迪FACS CantoⅡ流式细胞仪性能评价及实验室应用[J].现代检验医学杂志,2019,34(02):94.[doi:10.3969/j.issn.1671-7414.2019.02.025]
 FANG Pei-qi,SUN Lin,GU Mei-xiu,et al.Evaluation and Clinical Application of BD FACS Canto II Flow Cytometer[J].Journal of Modern Laboratory Medicine,2019,34(02):94.[doi:10.3969/j.issn.1671-7414.2019.02.025]

备注/Memo

备注/Memo:
作者简介:龙一飞(1986-),男,本科,初级技师,主要从事临床血液、体液学检验工作,Tel:13828490702,E-mail:long001fei@126.com。
更新日期/Last Update: 2015-03-20