[1]张莹,周铁成,童开,等.以LDH为例来探讨具有溯源性和互换性的校准品在临床生化应用中的重要性[J].现代检验医学杂志,2015,30(06):105-108.[doi:10.3969/j.issn.1671-7414.2015.06.032]
 ZHANG Ying,ZHOU Tie-cheng,TONG Kai,et al.With Lactate Dehydrogenase as An Example to Evaluate the Importance of the Traceability and Commutability Calibration in Clinical Biochemical Detection[J].Journal of Modern Laboratory Medicine,2015,30(06):105-108.[doi:10.3969/j.issn.1671-7414.2015.06.032]
点击复制

以LDH为例来探讨具有溯源性和互换性的校准品在临床生化应用中的重要性()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第30卷
期数:
2015年06期
页码:
105-108
栏目:
质量控制·实验室管理
出版日期:
2015-12-30

文章信息/Info

Title:
With Lactate Dehydrogenase as An Example to Evaluate the Importance of the Traceability and Commutability Calibration in Clinical Biochemical Detection
作者:
张莹周铁成童开马越云郝晓柯
第四军医大学西京医院全军临床检验中心,西安710032
Author(s):
ZHANG YingZHOU Tie-chengTONG KaiMA Yue-yunHAO Xiao-ke
Center of Clinic Laboratory Medicine of People’s Liberation Army, Xijing Hospital of Fourth Military Medical University,Xi’an 710032,China
关键词:
乳酸脱氢酶溯源性互换性校准品
分类号:
R446
DOI:
10.3969/j.issn.1671-7414.2015.06.032
文献标志码:
A
摘要:
目的通过在临床实验室使用具有溯源性和互换性的校准品,寻求血清酶测定标准化的新途径。方法以乳酸脱氢酶(LDH)为例,收集混合血清,由4家酶学参考实验室联合赋值,制备具有溯源性和互换性的参考物质。参加试验的8个实验室先使用自己的校准品,在校准通过后,将高中低3个水平的血清样本和参考物质各测定3次,记录检验结果;再使用具有溯源性和互换性的参考物质作为校准品校准后,重复测定高中低3个水平的血清样本和参考物质各3次,记录检验结果。计算8家实验室在使用具有溯源性和互换性的参考物质作为校准品校准前后,参考物质检测结果与靶值的偏倚和高中低3个水平的血清样本LDH检测结果的变异系数(CV%)。结果使用具有溯源性和互换性的参考物质参后,参考物质检测结果与靶值的偏倚由6.80%降到1.40%;高中低3个水平的血清样本LDH检测结果的变异系数CV由11.41%,9.93%和9.87%降到了4.13%,3.55%和3.80%。结论在临床实验室使用具有溯源性和互换性的参考物质作校准品,能大大降低实验室间结果的正确度和精密度,提高检测结果的准确性,有利于各个实验室各设备间结果的标准化和一致性。
Abstract:
ObjectiveWith the use of traceability and commutability calibration in clinical laboratory,try to seek new way in the standardized of serum enzyme determination.MethodsFor example to the lactate dehydrogenase (LDH),to collect mixed serum,which assigned by four enzymology reference laboratory,produced the traceability and commutability reference materials.Firstly,eight laboratories used their own calibration to calibrate their detection systems.After calibration,every laboratory detected one enzyme calibrator and three patient samples and then recorded the results.Additionally reused the traceability and commutability reference materials to calibrate their detection systems.After calibration,every laboratory detected them again and recorded the results.Calculated the calibration products for the determination of the bias between the results of the overall mean and labeled values,determination of serum samples from the changes of the results of precision before and after calibration.ResultsAfter the use of thetraceability and commutabilityreference materials,the bias between results of reference materials test and the target was value from 6.80% decline to 1.40%.The precision of three levels of serum samples were from 11.41%,9.93% and 9.93% to 4.13%,3.55% and 3.80%.ConclusionUsing the traceability and commutability reference materials as calibration in clinical laboratory,can greatly reduce the accuracy and precision of the results among these laboratory and improve the accuracy of test results,and then is increase the standardization and consistency in clinical laboratory.

参考文献/References:

[1]Miller WG, Myers GL .Commutability still matters[J].Clin Chem,2013,59(9):1291-1293.
[2]Zegers I,Beetham R,Keller T,et al.The importance of commutability of reference materials used as calibrators:the examples of ceruloplasmin[J].Clin Chem,2013,59(9):3122-1329.
[3]Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃.Part3.Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase[J].Clin Chem Lab Med,2002,40(16):642-648.
[4]张莹,周铁成,童开,等.酶学参考实验室室间质量评价技术方案的建立[J].现代检验医学杂志,2010,25(3):146-148. Zhang Y,Zhou TC,Tong K,et al.Established technical solutions of the enzymology reference laboratory extenal quality assessment[J].J Mod Lab Med,2010,25(3):146-148.
[5]International Organization for Stand ardization(ISO).In vitro diagnostic medical devices-measurement of quantitatias in biological samples-metrological traceability of values assigned to calibrators and control materials[S].ISO 17511,ISO,2003.
[6]Greg Miller W,Myers GL,Lou Gantzer M,et al.Roadmap for harmonization of clinical laboratory measurement procedures[J].Clin Chem,2011,57(8):1108-1117.
[7]Carobene A,Guerra E,Ceriotti F.A mechanism-based way to evaluate commutability of control materials for enzymatic measurements.The example of gamma-glutamyltransferase[J].Clinica Chimica Acta,2013(424):153-158.
[8]Perich C,Ricós C,Alvarez Z,et al.External quality assurance programs as a tool for verifying standardization of measurement procedures:Pilot collaboration in Europe[J].Clinica Chimica Acta,2014(432):82-89.
[9]李民,王延伟,陈阵,等.几种α-淀粉酶校准品和质控品的互换性研究[J].临床军医杂志,2012,40(4):942-943. Li M,Wang YW,Chen Z,et al.Evaluation of the commutability of several calibrators and quality control materials of the α-amylase[J].Clinical Journal of Medical Officers,2012,40(4):942-943.

相似文献/References:

[1]李大鹏,王永清,吕春雷,等.创伤性失血性休克大鼠血浆乳酸脱氢酶的动态测定及价值[J].现代检验医学杂志,2015,30(05):56.[doi:10.3969/j.issn.1671-7414.2015.05.017]
 LI Da-peng,WANG Yong-qing,L Chun-lei,et al.Dynamic Observation and Value of Plasma Level of Lactate Dehydrogenase in Rats with Traumatic Hemorrhagic Shock[J].Journal of Modern Laboratory Medicine,2015,30(06):56.[doi:10.3969/j.issn.1671-7414.2015.05.017]
[2]王伟伟,张 夏,黄传荣,等.血清LDH,TBIL,IBIL和Hcy水平检测 在MA和MDS临床鉴别诊断中的作用研究[J].现代检验医学杂志,2015,30(01):111.[doi:10.3969/j.issn.1671-7414.2015.01.031]
 WANG Wei-wei,ZHANG Xia,HUANG Chuan-rong,et al.Research on Detecting Serum LDH,TBIL,IBIL and Hcy Levels in Clinical Differential Diagnosis of MA and MDS[J].Journal of Modern Laboratory Medicine,2015,30(06):111.[doi:10.3969/j.issn.1671-7414.2015.01.031]
[3]游青青a,杨海健a,杨宇b,等.抑郁症患者血清脂质及CK,LDH水平与疾病程度的相关性分析[J].现代检验医学杂志,2019,34(01):79.[doi:10.3969/j.issn.1671-7414.2019.01.020]
 YOU Qing-qinga,YANG Hai-jiana,YANG Yub,et al.Correlation Analysis of Serum Lipids,CK,LDH and Course of Disease in Patients with Depression[J].Journal of Modern Laboratory Medicine,2019,34(06):79.[doi:10.3969/j.issn.1671-7414.2019.01.020]
[4]周莉莉a,刘 栋a,耿 丹a,等.陕西铜川地区< 18 岁健康人群血清CK,CK-MB,LDH 参考区间的建立与验证[J].现代检验医学杂志,2023,38(02):176.[doi:10.3969/j.issn.1671-7414.2023.02.033 ]
 ZHOU Li-lia,LIU Donga,GENG Dana,et al.Establishment and Verification of Reference Interval of Serum CK, CK-MB and LDH in Healthy Population Under 18 Years Old in Tongchuan Area of Shaanxi[J].Journal of Modern Laboratory Medicine,2023,38(06):176.[doi:10.3969/j.issn.1671-7414.2023.02.033 ]

备注/Memo

备注/Memo:
作者简介:张莹(1981-),女,硕士研究生,主管技师,主要从事临床生化及酶学参考实验室室间比对工作,Tel:13759901311,E-mail:xxfzzy@163.com。 通讯作者:周铁成,E-mail:tiecheng@fmmu.edu.cn。
更新日期/Last Update: 1900-01-01