[1]汪靖园,王林川,肖 尧,等.科华ELISA与雅培i2000sr检测系统对4种感染性疾病8项血清标志物检测的性能评价[J].现代检验医学杂志,2018,33(06):125-131.[doi:10.3969/j.issn.1671-7414.2018.06.033]
 WANG Jing-yuan,WANG Lin-chuan,XIAO Yao,et al.Performance Evaluation of Kehua ELISA and Abbott i2000sr in Eight Serum Markers Detection for Four Infectious Diseases[J].Journal of Modern Laboratory Medicine,2018,33(06):125-131.[doi:10.3969/j.issn.1671-7414.2018.06.033]
点击复制

科华ELISA与雅培i2000sr检测系统对4种感染性疾病8项血清标志物检测的性能评价()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第33卷
期数:
2018年06期
页码:
125-131
栏目:
研究简报·实验技术
出版日期:
2018-12-24

文章信息/Info

Title:
Performance Evaluation of Kehua ELISA and Abbott i2000sr in Eight Serum Markers Detection for Four Infectious Diseases
文章编号:
1671-7414(2018)06-125-07
作者:
汪靖园1王林川1肖 尧1阮竞雄1田旭东1闫 芳2
1.西安交通大学第一附属医院检验科,西安 710061; 2.西安市第三医院输血科,西安 710018
Author(s):
WANG Jing-yuan1WANG Lin-chuan1XIAO Yao1RUAN Jing-xiong1TIAN Xu-dong1YAN Fang2
1.Department of Clinical Laboratory,the First Affiliated Hospital of Xi'an Jiaotong University,Xi'an 710061,China; 2.Department of Blood Transfusion,Xi'an No.3 Hospital,Xi'an 710018,China
关键词:
科华ELISA 雅培i2000sr 血清标志物 性能评价
分类号:
R446
DOI:
10.3969/j.issn.1671-7414.2018.06.033
文献标志码:
A
摘要:
目的 对科华ELISA和雅培i2000sr检测系统4种感染性疾病8项血清学标志物的性能进行评价和比较。方法 采用美国临床和实验室标准协会(clinical and laboratory standards institute,CLSI)评价方案(evaluation protocols,EP)评价两种检测系统性能。评价内容包括:准确度、检出限、一致性、精密度、线性范围。结果 两种检测系统的准确度均为100%。HBsAg,HBsAb,HBeAg,HBeAb,HBcAb,Anti-HCV,梅毒特异性抗体及(Anti-TP)Anti-HIV的检出限科华分别为0.08IU/ml,10mIU/ml,0.8NCU/ml,4.4NCU/ml,3.2NCU/ml,0.8NCU/ml,20 mIU/ml和0.5 NCU/ml; 雅培分别为0.05 IU/ml,10 mIU/ml,0.23 NCU/ml,1 NCU/ml,1.8 NCU/ml,0.45 NCU/ml,20 mIU/ml和2.88 NCU/ml。一致性验证:两种系统方法学比对Kappa值分布于0.8~0.9,定性项目仪器间比对Kappa值均为1, 定量项目(HBsAg,HBsAb)仪器间比对观察误差<1/2TEa,浓度误差指数绝对值<1。精密度评价:批内变异系数(CV)范围为1.10%~7.01%,室内CV范围为1.90%~10.03%。雅培i2000sr HBsAg,HBsAb线性范围分别为0.73~245IU/ml和4.21~960.89mIU/ml。结论 科华ELISA和雅培i2000sr准确度、检出限、一致性、精密度和线性范围与厂家声明一致并能满足实验室要求。两种系统HBsAg,HBsAb,Anti-TP检出限接近,但HBeAg,HBeAb,HBcAb,Anti-HCV和Anti-HIV检出限二者存在较大差异。
Abstract:
Objective The present study was carried out toevaluate the performance of Kehua ELISA and Abbott i2000sr for eight serum markers detection.Methods The performances,i.e.,accuracy,limit of detection(LOD),consistency,precision and linear range were evaluated according with the CLSI EP documents.Results The accuracy of the two systems were 100%.The LOD of HBsAg,HBsAb,HBeAg,HBeAb,HBcAb,Anti-HCV,Anti-TP and Anti-HIV were 0.08IU/ml,10mIU/ml,0.8NCU/ml,4.4NCU/ml,3.2NCU/ml,0.8NCU/ml,20mIU/ml and 0.5NCU/ml for Kehua,and 0.05IU/ml,10mIU/ml,0.23NCU/ml,1NCU/ml,1.8NCU/ml,0.45NCU/ml,20mIU/ml and 2.88NCU/ml for Abbott,respectively.The consistency test showed that the Kappa value was 0.8 to 0.9 for method comparison and 1 for instruments comparison.Multipleinstruments comparison for HBsAg and HBsAb,the observed error was <1/2TEa anderror index by concentration was <1.The coefficients of variations(CV)ofwithin-run and laboratory precision were 1.10%~7.01% and 1.90~10.03%,respectively.The linear ranges of HBsAg and HBsAb were 0.73~245IU/ml and 4.21~960.89mIU/ml,respectively.Conclusion The performances oftwo systems,such as LOD,consistency,precision and linear range,were consistent with manufacturer's declaration and met the clinical requirements.Kehua was close to i2000sr on the detection ability of HBsAg,HBsAb and Anti-TP, however,obvious difference in the detection ability of HBeAg,HBeAb,HBcAb, Anti-HCV and anti-HIV was observed between two systems.

参考文献/References:

[1] 中国合格评定国家认可委员会.医学实验室质量和能力认可准则在临床免疫学定性检验领域的应用说明[S].北京:中国标准出版社,2012.
China National Accreditation Service for Conformity Assessment.Guidance on the application of accreditation criteria for the medical laboratory quality and competence in the field of clinical qualitative immunology[S].Beijing:China Standards Press,2012.
[2] Clinical and Laboratory Standards Institute.User Ve-rification of Precision and Estimation of Bias,3rd Edition[S].Wayne:PA,CLSI EP15-A3,2014.
[3] Clinical and Laboratory Standards Institute.Evaluation of the Linearityof Quantitative Measurement Procedures:A Statistical Approach,1st Edition[S].Wayne:PA,CLSI EP6-A,2003.
[4] Clinical and Laboratory Standards Institute.User Pro-tocol for Evaluation of Qualitative Test Performance,2nd Edition[S].Wayne:PA,CLSI EP12-A2,2008.
[5] 尚 红,王毓三,申子瑜.全国临床检验操作规程[S].北京:人民卫生出版社,2015:1016.
Shang H,Wang YS,Shen ZY.National guide to clinical laboratory procedures[S].4th Ed.Beijing:People's Medical Publishing House,2015:1016.
[6] 刘 杨,彭道荣,马越云,等.2种HBsAg定量检测系统对HBV不同感染阶段及不同基因型样本检测的一致性评价[J].检验医学,2016,31(04):288-292.
Liu Y,Peng DR,MA YY,et al.Consistency of 2 HBsAg quantitation determination systems in different HBV infection phases and different genotypes[J].Laboratory Medicine,2016,31(04):288-292.
[7] Fruttero ER,Picciolo G,De Matteis JA.Operative pe-rformances and efficiency for infectious disease testing of two immunochemistry analyzers-Abbott ARCHITECT i2000SR and DiaSorin Liaison XL[J].Clinical Chemistry & Laboratory Medicine,2015,53(1):e19-e23.
[8] 孙 彬,李 康,吴 纯,等.罗氏Cobas e601与雅培Architect i2000检测乙肝表面抗原的比较[J].国际检验医学杂志,2014,35(20):2823-2824,2851.
Sun B,Li K,Wu C,et al.Comparison of HBsAg results between RocheEobas e601 and Abbott Architect i2000 analysis systems[J].International Journal of Laboratory Medicine,2014,35(20):2823-2824,2851.
[9] 李 娅,章 迪,张 赟,等.罗氏cobas e602与modular e170电化学发光分析仪检测血清HBsAg阳性低值结果的对比分析[J].现代检验医学杂志,2017,32(3):123-125.
Li Y,Zhang D,Zhang Y,et al.Study on the low positive results ofdetecting HBsAg in serum specimens with Roche Cobas e602 and Modular e170 electrochemical luminescence analyzer[J].Journal of Modern Laboratory Medicine,2017,32(3):123-125.
[10] 邹 麟,张莉萍,夏吉荣,等.全自动免疫分析仪ARCHITECT i2000检测HBV血清标志物性能评价[J].重庆医学,2010,39(24):3353-3354.
Zou L,Zhang LP,Xia JR,et al.Efficiency evaluation of detecting HBV serum markers by ARCHITECT i2000 chemoluminescence analyzer[J].Chongqing Medicine,2010,39(24):3353-3354.
[11] 卢 婷,李发科,罗 杰,等.酶联免疫吸附试验检测乙型肝炎病毒血清学标志物的性能评价[J].检验医学与临床,2015,12(5):582-584.
Lu T,Li FK,Luo J,et al.Performance evaluation of ELISA for detecting HBV serological markers[J].Laboratory Medicine and Clinic,2015,12(5):582-584.
[12] 中国合格评定国家认可委员会.CNAS-RL02:能力验证规则[S].北京:中国标准出版社,2016.
China National Accreditation Service for Conformity Assessment.CNAS-RL02:Rules for Proficiency Testing[S].Beijing:China Standards Press,2016.

备注/Memo

备注/Memo:
基金项目:陕西省自然科学基金资助项目(No.2017JM8121)。
作者简介:汪靖园(1983-),女,硕士,主管技师,主要从事临床检验,研究方向:病原生物学的研究,E-mail:123436110@qq.com。
通讯作者:闫 芳(1973-),女,主管技师,主要从事临床检验,研究方向:临床用血安全,E-mail:173032147@qq.com。
更新日期/Last Update: 2018-11-30