[1]李寿敏,张 茜,赵 府,等.严重脂血对Sysmex XE-2100 全自动血液分析仪血细胞检测结果的影响[J].现代检验医学杂志,2020,35(02):100-103.[doi:10.3969/j.issn.1671-7414.2020.02.028]
 LI Shou-min,ZHANG Qian,ZHAO Fu,et al.Severel Pemia on Blood Routine Test Results ofSysmex XE-2100 Automatic Hematology Analyzer[J].Journal of Modern Laboratory Medicine,2020,35(02):100-103.[doi:10.3969/j.issn.1671-7414.2020.02.028]
点击复制

严重脂血对Sysmex XE-2100 全自动血液分析仪血细胞检测结果的影响()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第35卷
期数:
2020年02期
页码:
100-103
栏目:
研究简报·实验技术
出版日期:
2020-05-24

文章信息/Info

Title:
Severel Pemia on Blood Routine Test Results ofSysmex XE-2100 Automatic Hematology Analyzer
文章编号:
1671-7414(2020)02-100-04
作者:
李寿敏1张 茜1赵 府1郭子微2惠 维2王 诚1
(1. 六盘水市人民医院检验科,贵州六盘水 553001;2. 遵义医科大学,贵州遵义 563000)
Author(s):
LI Shou-min1ZHANG Qian1ZHAO Fu1GUO Zi-wei2HUI Wei2WANG Cheng1
(1.Department of Clinical Laboratory,Renmin Hospital of Liupanshui City, Guizhou Liupanshui 553001, China;2.Zunyi Medical University, Guizhou Zunyi 56300,China)
关键词:
脂血Sysmex XE-2100影响
分类号:
R446.111
DOI:
10.3969/j.issn.1671-7414.2020.02.028
文献标志码:
A
摘要:
目的 研究严重脂血对Sysmex XE-2100 全自动血液分析仪血细胞检测结果的影响。方法 将30 例无脂血、溶血、黄疸的标本进行血细胞检测及TG 检测作为对照组。30 例每份标本平均分两管,A,B 两组标本分别吸出70,120μl 血浆加入同等量的脂肪乳制备乳糜血。血细胞检测用Sysmex XE-2100 血液分析仪检测,TG 水平用Rochecobas-8000 生化分析仪检测,然后根据脂血程度分组将检测结果与对照组进行比较。结果 对于Sysmex XE-2100,A 组和B 组中Hb,MCH,MCHC,WBC,LYMPH%,MODO%,RDW-SD,RDW-CV 水平与对照组相比,差异均有统计学意义(t = -20.722~15.005,均P < 0.05)。B 组中HCT,MCV,NEUT% 水平与对照组相比,差异均有统计学意义(t= -5.829~-2.836,均P < 0.05)。RBC,PLT,EOS% 和BASO% 水平在两个组中与对照组相比,差异无统计学意义(t=-0.054~1.229,均P > 0.05)。结论 不同程度脂血会对Sysmex XE-2100 血液分析仪检测指标产生不同的影响,首先受到影响的指标Hb,MCH,MCHC,WBC,LYMPH%,MONO%,RDW-SD 和RDW-CV,当脂血程度达到更高时还会对HCT,MCV和NEUT% 造成影响,同时会造成白细胞不分类的情况,RBC,PLT,EOS% 和BASO% 结果在本实验中不受脂血影响。

参考文献/References:

[1] 周窈佳.脂血对Sysmex XE-5000 血液分析仪检测指标的影响[J].检验医学与临床,2014,11(2):206-207.  ZHOU Yaojia .Influences of lipemia on detectionindicator of Sysmex XE-5000 hematology analysis [J].Laboratory Medicine and Clinic, 2014,11(2):206-207.
[2] 徐玉兵,高春芳,赵琳.脂血对Sysmex XE-2100D血液分析仪检测指标的影响研究[J].检验医学,2012,27(12):1017-1020.   XU Yubing,GAO Chunfang,ZHAO Lin.Theresearch on the influence of lipemia on the detectionparameters of Sysmex XE-2100D hematology analyzer[J]. Laboratory Medicine,2012,27(12):1017-1020.
[3] 马骥,杨游萍,陈炜烨,等.高脂血对两种方法测定血红蛋白的影响及其校正[J].国际检验医学杂志,2012,33(9):1045-1046.   MA Ji,YANG Youping,CHEN Weiye,et al. Influenceof blood fat level on two methods for determination ofhemoglobin and its correction [J]. International Journalof Laboratory Medicine, 2012,33(9):1045-1046.
[4] 李劲,姚玉轩.重度脂血对Sysmex 全自动五分类血细胞分析仪检测结果的影响[J].淮海医药,2013,31(1):58-59.  LI Jin,YAO Yuxuan.Influence of severe lipemia ondetection of Sysmex automatic five parts hematologyanalyzers[J].Journal of Huaihai Medicine , 2013,31(1):58-59.
[5] 刘艳红,李南.中、重度脂血对Sysmex XE-2100 及XN-9000 血液分析仪检测指标的影响比较[J].检验医学与临床,2017,14(12):1735-1737, 1740.  LIU Yanhong,LI Nan.Comparison on influences ofmoderate and severe lipemia on detection indicatorsof Sysmex XE-2100 and XN-9000 hematologyanalyzers[J].Laboratory Medicine and Clinic,2017,14(12):1735-1737, 1740.
[6] 王克迪,徐东江,苏建荣.血浆置换法和公式校正法纠正乳糜血对仪器法测定血红蛋白影响的探讨[J]. 现代检验医学杂志,2017,32(3): 137-139,143.  WANG Kedi,XU Dongjiang,SU Jianrong. Correctionof the impact of chylemia on hematology analysis byplasma exchange and formula calibration method[J].Journal of Modern Laboratory Medicine,2017,32(3):137-139,143.
[7] 王利娜.脂血标本对实验室各常规检测项目的影响及处理方法[J].中国医学检验杂志,2012(3):99-100.  WANG Lina. The effect of lipemic specimens onlaboratory routine test items and the countermeasures[J].Chinese Journal of Medical Laboratory Technology,2012(3):99-100.
[8] 潘晓骅.重度脂血标本两种血红蛋白浓度校正方法的比较[J].国际检验医学杂志,2017,38(A02):162-163.  PAN Xiaohua.Comparison of two hemoglobinconcentration correction methods in severe lipidbooldsamples[J]. International Journal of LaboratoryMedicine,2017,38(A02):162-163.
[9] 刘成玉,罗春丽.临床检验基础[M].5 版. 北京:科学技术文献出版社,2017:34.   LIU Chenyu,LUO Chunli. Basis of clinicalexamination[M].5th Ed.Beijing: Science andTechnology Academic Press,2017:34.
[10] 冯惠,周世超.血液标本中不同浓度脂质对血常规检验结果的影响[J].中医临床研究,2018,10(15):49-50.   FENG Hui, ZHOU Shichao.Effects of differentconcentrations of lipids in blood samples on bloodtest reults[J]. Chinese Journal of Clinical Research ,2018,10(15):49-50.
[11] 张正云,何清.高脂血标本对血红蛋白的干扰及校正方法[J]. 实验与检验医学,2014,32(5):644-645.   ZHANG Zhengyun,HE Qing. Interference ofhyperlipidemia samples to hemoglobin and itscorrection method[J]. Experimental and LaboratoryMedicine, 2014,32(5):644-645.
[12] 曾素根,余江,曾婷婷,等.Sysmex 公司血液分析仪的干扰因素分析判断及处理程序[J].检验医学,2010,25(3):244-246.   ZENG Sugen,YU Jiang,ZENG Tingting, etal.Analysis and treatment program of interferencefactors of hematologyanalyser in SysmexCounpany[J],Laboratory Medicine , 2010,25(3):244-246.
[13] ZENG Sugen, ZENG Tingting, JIANG Hong, et al.A simple, fast correction method of triglycerideinterference in blood hemoglobin automatedmeasurement.[J] Journal of Clinical LaboratoryAnalysis,2013, 27(5) :341-345.

备注/Memo

备注/Memo:
作者简介: 李寿敏(1987-),男,本科,医学学士, 检验师,主要从事临床血液学研究,E-mail:378219117@qq.com。收稿日期:2019-10-01修回日期:2019-11-06
更新日期/Last Update: 2020-04-30