[1]孔丽蕊,黄 英,张 艳,等.血糖、糖化血红蛋白的参考变化值和参考变化因子的应用研究[J].现代检验医学杂志,2020,35(03):110-111.[doi:10.3969/j.issn.1671-7414.2020.03.029]
 KONG Li-rui,HUANG Ying,ZHANG Yan,et al.Application Study of Reference Change Value and Reference Change Factor of Blood Glucose and Glycated Hemoglobin[J].Journal of Modern Laboratory Medicine,2020,35(03):110-111.[doi:10.3969/j.issn.1671-7414.2020.03.029]
点击复制

血糖、糖化血红蛋白的参考变化值和参考变化因子的应用研究()
分享到:

《现代检验医学杂志》[ISSN:/CN:]

卷:
第35卷
期数:
2020年03期
页码:
110-111
栏目:
研究简报·实验技术
出版日期:
2020-07-20

文章信息/Info

Title:
Application Study of Reference Change Value and Reference Change Factor of Blood Glucose and Glycated Hemoglobin
文章编号:
1671-7414(2020)03-110-03
作者:
孔丽蕊黄 英张 艳徐晓华罗 宇潘跃承刘 翔
(成都市郫都区中医医院检验科,成都 611730)
Author(s):
KONG Li-rui HUANG Ying ZHANG Yan XU Xiao-hua LUO Yu PAN Yue-cheng LIU Xiang
(Department of Clinical Laboratory, Traditional Chinese Medicine Hospital of Pidu District, Chengdu 611730, China)
关键词:
参考变化值参考变化因子血糖糖化血红蛋白
分类号:
R446.112
DOI:
10.3969/j.issn.1671-7414.2020.03.029
摘要:
目的 探讨血糖(GLU)、糖化血红蛋白(HbA1c)的参考变化值(reference change value, RCV) 和参考变化 因子(reference change factor, RCF)在临床的应用。方法 根据RCV 计算公式:RCV=Z×21/2×( CV2 A + CVI 2 )1/2 和RCF 上、下限计算公式:RCFup=exp(Z×21/2×( CV2 A + CVI 2 )1/2/100)以及RCFdown=1/RCFup,分别计算GLU,HbA1c 的RCV 和RCF 在不同概率P<0.05,P<0.01 时单双侧值。结果 RCV 在P<0.05 时单双侧值分别为GLU(13.50%,16.04%)和 HbA1c (6.03%,7.16%);在P<0.01 时单双侧值分别为GLU(19.07%,21.12%) 和HbA1c(8.51%,9.42%)。RCF 在P<0.05 时单双侧的RCFup 和RCFdown 分别为GLU(1.14,0.87;1.17,0.85) 和HbA1c(1.06,0.94;1.07,0.93);P<0.01 时,RCF 单双侧的RCFup 和RCFdown 分别为GLU(1.21,0.83;1.24,0.81) 和HbA1c(1.09,0.92;1.10,0.91)。结论 RCV 在相同 概率下双侧值大于单侧值;P<0.05 的RCV 值小于P<0.01 的RCV 值。RCF 在相同概率下,单侧和双侧RCF 的上、下 限值变化不大;P<0.05 的RCF 上、下限值大于P<0.01 的RCF 上、下限值。建议临床实验室在监测糖尿病患者GLU, HbA1c 指标时,结合RCV 和RCF 判读检测结果。
Abstract:
Objective To investigate the clinical application of reference change value (RCV) and reference change factor (RCF) for blood glucose (GLU) and glycated hemoglobin (HbA1c). Methods According to RCV calculation formula: RCV = Z×21/2×( CV2 A + CVI 2 )1/2 and RCF upper and lower limit calculation formula: RCFup = exp (Z×21/2×( CV2 A + CVI 2 )1/2/100) and RCFdown = 1 / RCFup. Single and double-sided values of RCV and RCF of GLU and HbA1c were calculated respectively at different probability P <0.05, P <0.01. Results When P <0.05, the unilateral and bilateral values of RCV were GLU (13.50%, 16.04%) and HbAIc (6.03%, 7.16%). When P <0.01, the unilateral and bilateral values of RCV were GLU (19.07%, 21.12%) and HbA1c (8.51%, 9.42%).When P<0.05, the RCFup and RCFdown values of unilateral and bilateral RCF were GLU(1.14, 0.87 and 1.17, 0.85), HbAIc(1.06, 0.94 and 1.07, 0.93). When P<0.01, the RCFup and RCFdown values of unilateral and bilateral RCF were GLU(1.21, 0.83;1.24, 0.81), HbA1c(1.09, 0.92 and 1.10, 0.91). Conclusion The bilateral value of RCV was greater than the unilateral value under the same probability. The value of RCV at P<0.05 was less than its value at P <0.05.Under the same probability, the upper and lower limits of one and two sides of RCF dose not change much. The upper and lower limits of RCF with P<0.05 were greater than those with P<0.01. It was suggested that the clinical laboratory should combine RCV and RCF to interpret the test results when monitoring GLU and HBA1C indicators in diabetic patients.

参考文献/References:

[1] HARRIS E K and YASAKA T. On the calculation of a “reference change” for comparing two consecutive measurements[J]. Clin Chem, 1983, 29(1): 25-30.
[2] FRASER C G. Reference change values[J]. Clin Chem Lab Med, 2012, 50(5): 807-812.
[3]  PETERSEN P H, FRASER C G. Strategies to set global analytical quality specifications in laboratory medicine: 10 years on from the Stockholm consensus conference[J]. Accred Qual Assur,2010, 15: 323-330.
[4] LUND F, PETERSEN P H, FRASER C G, et al. Calculation of limits for significant unidirectional changes in two or more serial results of a biomarker based on a computer simulation model. Ann Clin Biochem, 2015, 52(2): 237-244.
[5] COSKUN A,SERTESER M,KARPUZOGLU H F, et al. How can we evaluate differences between serial measurements on the same sample? A new approach based on within-subject biological variation[J]. Clin Chem Lab Med,2017, 55(2): e44-e46.
[6] 赵海建,张传宝,王薇,等. 卫生行业标准WS/ T403-2012 在全国常规化学室间质量评价和室内质 控中的应用 [J] . 中华检验医学杂志,2014,37( 11 ): 866-869.    ZHAO Haijian, ZHANG Chuanbao, WANG Wei, et al. Application of two analytical specifications in the external quality assessment schemes and the external comparision of internal quality control [J] .Chinese Journal of Laboratory Medicine, 2014,37 (11): 866-869.
[7]  QI Zhihong, ZHANG Lin, CHEN Yu, et al. Biological variations of seven tumor markers[J]. Clin Chim Acta. 2015, 450: 233-236.

相似文献/References:

[1]张 艳,李鸿钏,刘 翔,等.甲状腺疾病应用参考变化因子评估连续监测甲状腺功能指标结果的价值[J].现代检验医学杂志,2019,34(05):135.[doi:10.3969/j.issn.1671-7414.2019.05.034]
 ZHANG Yan,LI Hong-chuan,LIU Xiang,et al.Value of Using Reference Change Factor to Evaluate the Results of Continuous Monitoring of Thyroid Function Indicators in Thyroid Diseases[J].Journal of Modern Laboratory Medicine,2019,34(03):135.[doi:10.3969/j.issn.1671-7414.2019.05.034]

备注/Memo

备注/Memo:
收稿日期:2019-08-10 修回日期:2020-02-13作者简介:孔丽蕊(1972-),女,本科,副主任技师,研究方向:临床生物化学和免疫学,E-mail:524392472@qq.com
更新日期/Last Update: 2020-07-20